Global search

Find data and evidence

Type at least 2 characters. Use arrow keys to review and Enter to open.

Peer reviewedOpen accessNo known noticesLeptospirosis

Rapid and sensitive point-of-care detection of Leptospira by RPA-CRISPR/Cas12a targeting lipL32

PLOS Neglected Tropical Diseases ·

Sirawit Jirawannaporn, Umaporn Limothai, Sasipha Tachaboon, Janejira Dinhuzen, Patcharakorn Kiatamornrak, Watchadaporn Chaisuriyong, Jom Bhumitrakul, Oraphan Mayuramart, Sunchai Payungporn, Nattachai Srisawat

DOI
10.1371/journal.pntd.0010112
PMID
34990457
PMCID
PMC8769300
OpenAlex
W4206227220
Integrity status
No known notices
Study design
Journal article
Publisher
Public Library of Science (PLoS)
Article type
journal-article

The paper addresses the stated diagnostic gap of rapid, sensitive near-patient testing for leptospirosis. It presents a highly specific assay and a simpler readout format as a possible screening tool for resource-limited settings.

01

Structured evidence summary

Research question

The stated aim was to create and validate a CRISPR/Cas12a assay paired with isothermal amplification for detecting leptospires in extracted patient DNA, using lipL32 as the target. The abstract also describes a lateral-flow adaptation intended to make interpretation easier.

Study design

The report describes development of an RPA-CRISPR/Cas12a fluorescence assay targeting lipL32, analytical testing for detection threshold and cross-reactivity, and comparison with qPCR on DNA from 110 clinical specimens. A lateral-flow version was then evaluated in a pilot using known positive and negative samples.

Population and setting

The evaluations used extracted patient DNA and 110 clinical specimens from the reported clinical validation; the disease context is leptospirosis and diagnostics. The intended context proposed by the authors is acute leptospirosis screening in resource-limited settings; no country is listed in the supplied metadata.

Main findings

The assay detected Leptospira at 100 cells/mL and showed no observed cross-reactivity with several other acute febrile illnesses in the reported tests. In 110 clinical specimens, sensitivity was 85.2%, specificity 100%, and accuracy 92.7% versus qPCR; sensitivity was higher on days 4–6 after fever onset and lower after day 7. A lateral-flow version had the same detection threshold and distinguished known positive and negative clinical samples in a pilot; overall platform performance was reported as better than a commercial rapid test.

Public-health relevance

The paper addresses the stated diagnostic gap of rapid, sensitive near-patient testing for leptospirosis. It presents a highly specific assay and a simpler readout format as a possible screening tool for resource-limited settings.

Important limitations

No formal limitations are stated in the supplied abstract. The described evidence is an assay validation using 110 specimens plus a pilot lateral-flow evaluation; this single-article abstract/metadata summary requires the original paper for decision-grade interpretation.

GIDS interpretation

For GIDS discoverability, the supplied metadata places this as a peer-reviewed diagnostics article on leptospirosis; no country is listed. It provides context on a diagnostic assay only; this summary does not connect the article to any live surveillance signal.

02

Related GIDS surveillance

Literature links provide context, not evidence of a signal’s cause. Disease-only matches do not establish geographic relevance.

03

Evidence relationships

  • Disease Leptospirosis
  • Topic Diagnostics
  • Related Diagnostics
  • Pathogen type Bacterium
  • Study setting The evaluations used extracted patient DNA and 110 clinical specimens from the reported clinical validation; the disease context is leptospirosis and diagnostics. The intended context proposed by the authors is acute leptospirosis screening
  • Study design Journal article