Global search

Find data and evidence

Type at least 2 characters. Use arrow keys to review and Enter to open.

Peer reviewedOpen accessNo known noticesTuberculosis

Development of a Rapid Automated Point-of-Care Test for Mycobacterium tuberculosis Detection from Tongue Swabs and Sputum Specimens on the DASH® Rapid PCR System

Open Forum Infectious Diseases ·

Matthew A Butzler, Jennifer L Reed, Alaina M Olson, Rachel C Wood, Gerard A Cangelosi, Angelique K Luabeya, Mark Hatherill, Arthur M Chiwaya, Loren Rockman, Grant Theron, Sally M McFall

DOI
10.1093/ofid/ofag706
PMID
—
PMCID
—
OpenAlex
—
Integrity status
No known notices
Study design
Journal article
Publisher
Oxford University Press (OUP)
Article type
journal-article

Deploying rapid, near-care diagnostic tools in peripheral healthcare settings can facilitate earlier treatment initiation and enhance outcomes for tuberculosis globally.

01

Structured evidence summary

Research question

The study evaluated the performance and analytical metrics of a newly developed rapid polymerase chain reaction assay for detecting Mycobacterium tuberculosis using tongue swab and sputum samples.

Study design

This evaluation involved analytical testing for specificity and limit of detection alongside two clinical evaluations utilizing blinded, de-identified clinical specimens.

Population and setting

The research used specimens obtained from symptomatic clinic attendees located in South Africa.

Main findings

The assay achieved a sensitivity of 91% and specificity of 100% for tongue swabs, and a sensitivity of 96% with an 88% specificity for sputum specimens. Additionally, the system demonstrated cartridge stability and battery compatibility suited for resource-constrained environments.

Public-health relevance

Deploying rapid, near-care diagnostic tools in peripheral healthcare settings can facilitate earlier treatment initiation and enhance outcomes for tuberculosis globally.

Important limitations

Explicit limitations were not provided in the source text; therefore, the summary relies on the supplied single-article metadata and requires consultation of the original paper for decision-grade interpretation.

GIDS interpretation

This literature record describes diagnostic assay development and performance metrics in defined clinical evaluation groups, providing context on technological discoverability without establishing a live surveillance signal.

02

Related GIDS surveillance

Literature links provide context, not evidence of a signal’s cause. Disease-only matches do not establish geographic relevance.

03

Evidence relationships

  • Disease Tuberculosis
  • Topic Diagnostics
  • Topic Treatment
  • Related Diagnostics
  • Related Treatment
  • Study setting The research used specimens obtained from symptomatic clinic attendees located in South Africa.
  • Study design Journal article