Global search

Find data and evidence

Type at least 2 characters. Use arrow keys to review and Enter to open.

Peer reviewedOpen accessNo known noticesHepatitis BAcute hepatitis B

Nosocomial Transmission of Hepatitis B Virus Associated with the Use of a Spring-Loaded Finger-Stick Device

New England Journal of Medicine ·

Louis B. Polish, Craig N. Shapiro, Frederick Bauer, Phyllis Klotz, Paulette Ginier, Ronald R. Roberto, Harold S. Margolis, Miriam J. Alter

DOI
10.1056/nejm199203123261101
PMID
1738376
PMCID
—
OpenAlex
W2337624352
Integrity status
No known notices
Study design
Cohort study
Publisher
Massachusetts Medical Society
Article type
journal-article

The findings support careful use of finger-stick devices and strict adherence to precautions against blood contamination as measures that may reduce transmission of blood-borne pathogens in hospitals.

01

Structured evidence summary

Research question

The study investigated the source of acute hepatitis B virus infections clustered among hospitalized patients with diabetes, focusing on finger-stick blood sampling with a spring-loaded lancet device.

Study design

The investigators used a retrospective cohort study of patients with diabetes and a case-control study of nondiabetic patients and controls. They also reviewed relevant nursing procedures.

Population and setting

The setting was a single California hospital ward where 26 patients developed acute hepatitis B between June 1989 and March 1990; 23 had diabetes. The cohort comprised 72 patients with diabetes admitted during 1989, while the case-control comparison included three nondiabetic cases and 20 nondiabetic controls.

Main findings

Finger-stick sampling was associated with higher observed HBV attack rates among patients with diabetes (42% versus 0%; P=0.08) and with a dose-response pattern according to the number of finger sticks (P=0.002). All three nondiabetic cases had undergone finger-stick sampling, compared with none of the controls (P=0.0006); procedure review suggested that device-platform contamination was the likely transmission mechanism.

Public-health relevance

The findings support careful use of finger-stick devices and strict adherence to precautions against blood contamination as measures that may reduce transmission of blood-borne pathogens in hospitals.

Important limitations

The evidence is limited to the supplied single-article abstract and metadata, which do not provide enough methodological detail for decision-grade interpretation. The nondiabetic case-control comparison was small, and the retrospective cohort finding for the overall attack-rate difference was not statistically significant at the reported threshold.

GIDS interpretation

For discoverability, the paper is relevant to hepatitis B transmission dynamics and nosocomial infection associated with reusable finger-stick equipment. The supplied record is a peer-reviewed journal article; it does not constitute confirmation from a live surveillance system.

02

Related GIDS surveillance

Literature links provide context, not evidence of a signal’s cause. Disease-only matches do not establish geographic relevance.

03

Evidence relationships

  • Disease Hepatitis B
  • Topic Transmission dynamics
  • Pathogen type Virus
  • Study setting The setting was a single California hospital ward where 26 patients developed acute hepatitis B between June 1989 and March 1990; 23 had diabetes. The cohort comprised 72 patients with diabetes admitted during 1989, while the case-control c
  • Study design Cohort study